This article is a practical guide to Size Exclusion Chromatography covering instrumentation, calibration strategies, the data that can be extracted from a properly configured system, and the method development steps required to generate reliable and reproducible data. It assumes that the reader is already familiar with the basic principles of the technique. If you are new to SEC, we recommend reading our Introduction to GPC/SEC first, where the fundamental concepts are explained.
Size Exclusion Chromatography (SEC) – also called Gel Permeation Chromatography (GPC) – is a chromatographic technique that sorts molecules according to their size in solution or, in more technical terms, their hydrodynamic volume (Vh), the effective space a polymer coil occupies in solution. This sorting is done inside a column, where molecules are selectively retained to different degrees by the different pore sizes of the column’s packing.
SEC is the reference technique for measuring molecular weight distribution (MWD), averages (Mn, Mw, Mz), and dispersity, but because separation is done by hydrodynamic volume and not molecular weight, a calibration must be applied in order to obtain meaningful MWD data.
SEC columns contain a porous stationary phase, which is typically made of cross-linked polymer beads with a carefully engineered distribution of pore sizes. When a dissolved polymer sample passes through these columns, large molecules (those with large hydrodynamic volumes) cannot access the smaller pores and are carried rapidly through the interstitial volume (V0) between the particles, eluting first. Conversely, smaller molecules penetrate more of the total permeation volume (Vt), spending more time inside the stationary phase and eluting later.
This is the opposite of most other chromatographic techniques, where smaller analytes travel faster. In SEC, larger molecules elute first because the pore size distribution of the stationary phase controls which fraction of the total permeation volume is accessible to any given molecule.
Three terms describe what is mechanistically the same separation technique:
The mechanism – separating molecules by hydrodynamic volume using a porous stationary phase – is identical in all three cases. Throughout this article, both SEC and GPC will be used interchangeably.

Read more: GPC vs SEC vs GFC: Understanding the Difference & When to Use Each Term
A GPC/SEC analysis follows a simple sequence: pump → injector → column(s) → detector(s). Understanding what each component does, and where it can introduce errors, is what allows a polymer characterization laboratory to make sound decisions about setup, maintenance, and troubleshooting.
Solvent flow-rate stability directly determines retention time consistency. Any drift in pump delivery would directly translate into apparent shifts in molecular weight averages between injections. This is one of the reasons why pump precision is a non-negotiable specification in GPC/SEC instruments.
Retention time reproducibility is monitored in practice using a flow rate marker: a small, chemically inert molecule injected with each sample and whose elution time serves as an internal reference. Any shift in the marker’s elution time flags a change in effective flow rate.
The injector introduces a precise volume of polymer solution into the mobile phase stream. In automated GPC/SEC systems this is performed by an automated valve injector, which improves volume precision and run-to-run repeatability compared to manual injection.
Injection volume affects both peak shape and column loading, and should be optimized during method development. Overloading the column by injecting a volume too large or a solution too concentrated causes peak broadening and distorts molecular weight averages.
Every column system has two hard boundaries that define the useful analytical window:
The useful fractionation window sits between these two limits, and column selection is an exercise in matching this window to the expected molecular weight range of your sample. For instance, a column with predominantly large pores will poorly resolve the low-MW end, where species fully permeate regardless of size; and a column whose pores are too small relative to the sample’s MW range will push the high-MW fraction above the exclusion limit, where it co-elutes at the void volume with no recoverable resolution.
Three different parameters govern column performance:
Detector choice determines what information can be extracted from the chromatogram. The table below summarizes the main detector types used in GPC/SEC:
| Detector | Measures | Key Notes |
| IR (Infrared) | Concentration + chemical composition | Detects total C-H, CH2, CH3, and C=O groups. Limited by solvent IR overlap. |
| UV | Chromophore-containing polymers | Limited to polymers with UV-absorbing groups. |
| ELSD (Evaporative Light Scattering Detector) | Concentration | Universal signal. Highly sensitive, gradient-compatible, destructive detector, non-linear response, affected by different analytical variables |
| MALS (Multi-Angle Light Scattering) | Absolute Mw, Rg | Photodiodes placed at nominal angles from 20º to 150º Extrapolation to zero-angle Long-chain branching detection. |
| LALS, Low Angle Light Scattering | Absolute Mw | Photodiodes located at an angle low enough to consider the distance to zero angle negligible. No extrapolation required. No information on size. |
| Online Viscometer | Intrinsic viscosity [η] | Enables universal calibration and long-chain branching detection. |
A few important caveats that the table above cannot capture:
Light Scattering (LS) requires knowledge of the specific refractive index increment (dn/dc) for the polymer-solvent system at the measurement temperature. This value is not always available for novel polymers or for copolymers and must be measured independently. LS is a powerful absolute method, but it is not plug-and-play.
For polyolefin analysis, infrared detection provides simultaneous molecular weight distribution and chemical composition (short-chain branches / comonomer information) in a single analysis. This additional dimension can identify changes in molecular architecture that the MWD is blind to. This is data that neither Refractive Index nor Light Scattering can deliver.
Related article: Chemical Composition measurement in GPC/SEC analysis. Differentiating polyolefins with similar Molar Mass Distribution
Related article: How to choose the most appropriate combination of detectors for high temperature SEC/GPC analysis of polyolefins

Full SEC system and its components
Calibration is the intellectual core of SEC practice, and understanding the limitations of each type of calibration is key to obtaining reliable data.
The most widely used calibration approach uses a series of narrow-distribution standards to build a calibration curve of log(MW) versus elution volume. This curve relates a measured elution volume to a peak molecular weight for the standard polymer under the specific operating conditions.
Typical standard materials are Polystyrene (PS) and polymethyl methacrylate (PMMA) in organic systems, and Dextran and Pullulan in aqueous systems.
When, for instance, Polystyrene calibration is applied to characterize polyethylene (PE), a chemically and architecturally different polymer, the calibration curve maps the elution volume to a PS-equivalent MW, not the actual MW of the PE chain.
Conventional calibration is widely used due to its simplicity, but it assumes that the test material behaves similarly to the calibration standards in solution, and that differences in molecular architecture do not significantly affect hydrodynamic size.
There are established approaches to convert molecular weights relative to polystyrene (PS) standards into molecular weights relative to the polymer of interest. The Q-factor method is based on a vertical shift of the calibration curve, while the use of the Mark–Houwink coefficients of the polymer of interest provides an alternative approach.
Although less common nowadays, calibration methods based on broad standards are also available. In such cases, the reference materials are typically selected to match the chemistry of the samples being analyzed.
The solution to the limitations of conventional calibration was established by Benoit and co-workers in 1967 [1]. They observed that the product of intrinsic viscosity and molecular weight (proportional to the hydrodynamic volume) is invariant in SEC separation. For flexible-chain polymers – the vast majority of synthetic polymers – this holds regardless of polymer chemistry, chain architecture, or branching, provided the polymer is in the same solvent and at the same temperature.
This enables the application of universal calibration: a single calibration curve of log([η]·M) versus elution volume can be applied to any polymer.
The way to apply universal calibration in a GPC/SEC system is by combining a four-capillary, differential viscometer detector with a concentration detector (Refractive Index or Infrared). This way, intrinsic viscosity [η] is measured at every elution slice.
When analyzing polymers at high temperatures, it is important to consider that the hydrodynamic volume of a polymer coil at high temperature is not the same as it is at ambient temperature. Thermal expansion of the chain affects its solution conformation. This is one reason why Mark-Houwink parameters are temperature-specific, and why transferring ambient-temperature calibrations to HT-GPC conditions is not valid.
When a Light Scattering detector is combined with a concentration detector, the absolute weight-average molecular weight (Mw) at each elution slice is obtained directly from the excess Rayleigh scattering, without the need for a calibration curve [2]. MALS also yields the radius of gyration (Rg) as a function of MW, which is useful for long-chain branching analysis.
Absolute does not mean “assumption-free”. Reliable LS measurements require accurate dn/dc for the polymer-solvent system at the measurement temperature, accurate concentration at every slice of the chromatogram, and adequate signal-to-noise at the high- and low-MW tails (which are concentration-limited).
Related article: Calibration and Data Interpretation – Additional Guidance and Extended Best Practices
GPC/SEC characterization of polymers operates at two levels:
The first level (molecular weight distribution, averages, and dispersity) is accessible from any properly calibrated SEC system with a concentration detector, and it answers the most fundamental question in polymer characterization: how large are the polymer chains, and how broadly are they distributed?
The second level requires a SEC system with multi-detector configuration, and it answers the question that MW data alone cannot answer: what is the architecture of those chains?

GPC/SEC returns a distribution of molecular weight values, and it is from that distribution that a set of statistical averages can be calculated. Each average is a different mathematical moment of the distribution curve, weighted toward a different region of it.
| Average | Statistical Moment | Sensitive to | Structure-property relationship (Polyolefins) |
| Mn | First (number) | Low-MW tail: chain ends, oligomers, additives | Brittleness, solvent resistance, oligomers, extractables, migration (food-contact) |
| Mw | Second (weight) | Mid-distribution; the most commonly specified commercial parameter | Stiffness, tensile strength melt viscosity, resistance to slow crack growth |
| Mz / Mz+1 | Third / fourth | High-MW tail – vulnerable to truncation at the column exclusion limit | Melt elasticity, die swell, sagging in film and pipe extrusion, susceptibility to flow-induced crystallization |
| Ð = Mw/Mn | Ratio | Breadth of distribution (1.0 = monodisperse, theoretical limit) | Processing window width, property uniformity across a batch |
The practical significance of each average follows directly from which part of the distribution it reflects.
Read more: Molecular Weight Distribution in Polymers: Averages, Dispersity, and How to Measure Them
The structure-property relationships above are specific to polyolefins and are not transferable between polymer classes. Even PE and PP with identical Mw and Ð will not behave identically: crystallinity, chain stiffness, and the relationship between chain dimensions and mechanical response differ fundamentally between the two. Any structure-property relationships must be built from experimental data for each specific material system.
Molecular weight distribution describes chain size but does not provide information on chain architecture. Two polyethylene samples can exhibit identical molecular weight distributions and still behave very differently in processing and end use if one contains long-chain branching while the other does not, as branched molecules elute at higher elution volumes and therefore appear to have a lower apparent molecular weight under conventional calibration. Differences in comonomer distribution across the molecular weight range can also lead to significant variations in properties.
These are aspects that require a multi-detector configuration to address them.
The presence of long or short-chain branches is characteristic of some commercial polyolefin grades and, in some cases, both types of branches can be found in the same material. Long-chain and short-chain branching arise from different polymerization mechanisms, affect different material properties, and require different analytical approaches to be detected and quantified.
Long-chain branching refers to polymer branches of lengths that are a significant fraction of the length of the main polymer backbone. They arise from chain transfer to polymer during synthesis, or from specific catalytic mechanisms.
LCB impacts the rheological properties of polymers, such as the melt strength and stability, as well as the shear strength and elastic behavior in melt. Even low levels of LCB can produce significant changes in melt behavior.
A SEC system with a concentration detector (Infrared or Refractive Index) and additionally a viscometer or/and Light Scattering detector resolves LCB through the following logic: a branched chain occupies a smaller hydrodynamic volume than a linear chain at the same molecular weight. Therefore, LCB can be detected by comparing the size (through Rg) or compactness (through [h]) of the branched sample to that of the linear reference at the same molecular weight.
Related application note: Why Is My HDPE Failing In Production When MFI and Density Look Fine.
Both g and g’ are indexes sensitive to LCB: g=1 indicates lack of branching, and g<1 indicates presence of long-chain branches, the lower the value the higher the LCB content. The relationship between g and g′ is given by the viscosity shielding ratio: g′=gε. The value of ε is dependent on a number of factors, including solvent, temperature, and branching. ε has been found to generally fall in the range 0.5 to 1.5.
The parameters g and g′, both decrease with increasing degree of branching and can be used to calculate the number of branches per chain (Branching number: Bn) as a continuous function of the molar mass of the branched macromolecule, according to Flory’s theory. For polyolefins the branching points can be considered as trifunctional and in SEC monodisperse slices can be assumed and therefore the following equation is used to quantify Bn from g:

An additional useful parameter that can be calculated using the branching number is the branching frequency, λ, defined as the number of long-chain branches per 1000 repeat units, which can be calculated from Bn:

The Conformation Plot method is considered most accurate, based directly on Flory’s theory, although the Mark Houwink Plot has the advantage of higher precision, covering larger molar mass range (it can be measured to lower molar masses), and being less sensitive to non-ideal SEC behavior (as the late elution of highly branched high molar mass molecules which distort heavily the Conformation Plot). The drawback of the Mark-Houwink Plot method is that it requires assuming a value for the viscosity shielding ratio, ε, which may even depend on molar mass.
Short-chain branching refers to a macromolecular property distribution where small side branches are attached to the main polymer backbone. SCB occurs frequently in synthetic polymers like polyolefins, and it results from the incorporation of α-olefin comonomers into the polyethylene backbone during copolymerization.
The most common comonomers in commercial grades are 1-butene, 1-hexene, and 1-octene, producing pendant branches of 2, 4, and 6 carbons respectively. Each comonomer insertion is one branch.
Comonomer content and its distribution across the MW range are among the most consequential parameters in polyolefin characterization. SCB content and distribution directly controls crystallinity and crystallization kinetics: higher comonomer incorporation disrupts chain packing, reducing crystallinity and producing softer, more flexible material. Comonomer content also governs seal initiation temperature in packaging films, and environmental stress crack resistance (ESCR), which determines long-term performance in pipe and container applications.
In polypropylene, the methyl substituent on every other backbone carbon is itself an intrinsic form of SCB, and its stereoregularity – isotactic, syndiotactic, or atactic – determines crystallinity and mechanical performance.
Bulk comonomer content – provided by NMR or FTIR – doesn’t tell the full story. What matters is where in the molecular weight distribution the SCB resides. For example, in Ziegler-Natta LLDPE, comonomer is typically incorporated into shorter chains (the low-MW fraction carries more SCB than the high-MW fraction). In metallocene (single site) LLDPE, comonomer distributes far more uniformly across the MW range. Two materials can have identical bulk comonomer content and identical density and yet perform completely differently in film or pipe service because of this difference in how SCB is distributed across the MWD.
This is the measurement that standard GPC/SEC detectors cannot make. Refractive Index, Light Scattering, and Viscometry all respond to concentration or to the molar mass or hydrodynamic volume of the chain, but none carries information about the existence of pendant groups. An Infrared (IR) detector, however, responds to specific C-H absorption bands. The ratio of the methyl (CH₃) band intensity to the methylene (CH₂) band intensity, measured continuously across the elution profile, provides a quantitative measure of comonomer content at each MW slice.
The infrared detector at the core of Polymer Char’s GPC-IR instrument is designed for this measurement. The IR collects two absorbance channels simultaneously: the CH₂ band sensitive mostly to the polymer backbone carbons, and the CH₃ band, carrying the comonomer content signal. The ratio of these two signals, calibrated against standards of known comonomer content, yields short-chain branching frequency in branches per 1000 carbons as a continuous function of molecular weight – the SCB distribution across the full MWD – from a single injection.
Related article: Short-chain Branching in polyolefins: what causes short-chain branches and how to study them
Solvent and temperature selection is driven by polymer solubility, because a polymer that is not fully dissolved cannot be characterized.
Read more: Practical considerations in Gel Permeation Chromatography analysis of polyolefins
Shear degradation is a phenomenon in which high-MW polymer chains are degraded by shear forces, resulting in chain scission. In an SEC chromatogram this leads to a shift of the polymer peak towards greater elution volumes (lower molar masses).
Shear degradation can occur at these points:
Related Poster: Improvements in Sample Preparation of Polyolefins to prevent Polymer Degradation prior to GPC/SEC and CEF analysis
Related Poster: GPC analysis at different flow rates to overcome shear degradation
Filtration is required to protect the column from fillers, gels, or undissolved material.

The integration limits define what molecular weight range is included in the reported averages.
Poorly placed baseline and integration limits can inflate or truncate both the low-MW tail, directly distorting Mn (highly sensitive to low-MW content), and the high-MW tail, distorting Mz (sensitive to the highest-MW species). Baseline placement decisions should be documented and applied consistently across a sample set.
Method validation is the systematic process of demonstrating that the complete analytical system – instrument, column set, mobile phase, sample preparation protocol, detectors, and data processing parameters – delivers accurate, reproducible results for the specific materials being characterized.
a) System suitability and instrument qualification
System suitability checks should be run at the start of each analytical session to detect any instrument drift, column degradation, or solvent quality issues before they corrupt sample data. These routine operational controls should include:
b) Calibration verification
The calibration must be verified against reference materials before any sample series is run.
Calibration should be repeated whenever columns are replaced, mobile phase batches change, or the system is moved or serviced. The frequency of routine recalibration checks should be defined as part of the method and justified based on demonstrated system stability.
c) Precision and reproducibility
Method precision is established by repeat injections of the same sample under identical conditions (repeatability) and by measurements across different days, operators, or instruments (intermediate precision). For most polymer characterization applications, Mw repeatability of 2-5% RSD and Mn repeatability of 2-7% RSD across a minimum of five injections are reasonable targets, though acceptance criteria should be defined based on the end-use requirements of the data.
Mz and Mz+1 are more sensitive to integration limits and high-MW tail noise and will show higher variability; this should be documented and understood before these averages are used in specifications.
d) Artifact detection and system checks
A validated method includes defined checks for the most common sources of artifacts:
e) Documentation and traceability
A working methodology is only as reliable as its documentation. Molecular weight data are only reproducible when the full analytical context is reproducible.
The method record should specify, at minimum:
Any deviation from these parameters should be recorded.
Related article: High Temperature GPC (HT-GPC): A Complete Guide
Hydrodynamic volume (Vh)
The effective volume occupied by a polymer coil in solution, including the solvent molecules trapped within and around the coil as it moves through the liquid. This property depends on the polymer-solvent interaction, temperature, and chain architecture. A polymer in a good solvent will swell and occupy a larger hydrodynamic volume than the same chain in a poor solvent. Branched chains occupy a smaller hydrodynamic volume than linear chains of the same molecular weight. Hydrodynamic volume is the property that SEC actually separates by, which is why two polymers of identical molecular weight but different architecture or chemistry can elute at different volumes.
Interstitial volume (V0)
Also referred to as void volume or dead volume.
The volume of mobile phase occupying the spaces between the packing particles in the column, excluding the pore interior. Molecules too large to enter any pore travel exclusively through this space and elute first, all at the same elution volume regardless of their actual size. It represents the upper boundary of the SEC separation window: any material eluting at or before V0 carries no molecular weight information.
Total permeation volume (Vt)
The total volume of mobile phase accessible to the smallest molecules in the system (the sum of the interstitial volume and the entire pore volume of the stationary phase). Molecules small enough to fully penetrate all pores elute at Vt, again all at the same volume regardless of size. It represents the lower boundary of the SEC separation window. The useful fractionation range lies between V0 and Vt.
Elution volume (Ve)
The volume of mobile phase that has passed through the column at the point when a given molecule or polymer fraction exits the column and reaches the detector. It is the primary experimental observable in SEC, from which molecular weight is calculated via the calibration curve. Elution volume is directly dependent on flow rate, temperature, and column condition, which is why these parameters must be tightly controlled and monitored throughout an analysis.
Theoretical plates / plate count (N)
A measure of column separation efficiency. A high plate count means the column produces sharp, narrow peaks and can resolve polymer fractions that differ only slightly in hydrodynamic volume. In practice, plate count is determined by injecting a narrow-distribution standard and measuring the width of the resulting peak relative to its elution volume. It is directly affected by particle size – smaller packing particles generate higher plate counts – and degrades over time as the column bed compresses or contaminates. Monitoring plate count regularly is one of the primary indicators of column health.
dn/dc (specific refractive index increment)
The change in refractive index of a polymer solution with polymer concentration, at constant temperature and solvent. It is a physical property specific to each polymer-solvent-temperature combination and a required input for any LS-based absolute molecular weight calculation.
It can be measured offline using a differential refractometer, or estimated online from the RI detector signal under the assumption that 100% of the injected mass is recovered from the column.
Intrinsic viscosity ([η])
A measure of the contribution of a single polymer chain to the viscosity of a solution, independent of concentration. It reflects the hydrodynamic volume of the chain in a given solvent at a given temperature: a larger, more expanded coil increases solution viscosity more than a compact or branched one of the same molecular weight. Intrinsic viscosity has units of dL/g and is related to molecular weight through the Mark-Houwink equation. In SEC, it is measured continuously across the elution profile by an online differential viscometer, providing a hydrodynamic size value at every MW slice.
Band broadening (BB)
Also referred to as peak broadening or zone broadening.
The spreading of a polymer peak as it travels through the SEC system. In an ideal separation, a monodisperse sample would elute as a perfectly sharp peak; in practice, the peak is always broader than the true sample distribution. This instrumental artifact causes the measured MWD to appear wider than it actually is, overestimating dispersity and distorting Mn and Mw.
Main causes: axial diffusion of molecules along the flow direction; mass transfer resistance between mobile phase and pore interior; non-uniform column packing; and extra-column volume from tubing, injector, and detector cells.
Mass recovery (%MR)
The ratio of the polymer mass detected by the concentration detector to the mass entered by the operator at the time of injection, expressed as a percentage. Values close to 100% indicate that the detected concentration matches the prepared concentration. Values below 100% indicate that a portion of the sample was not detected, most commonly due to incomplete dissolution, with undissolved material either retained on the in-line filter or settled out of solution before injection. Mass recovery is therefore a direct and quantitative check on dissolution quality, and more reliable than visual inspection of the solution alone.
[1] Grubisic, Z.; Rempp, P.; Benoit, H. J. Polym. Sci. B: Polym. Lett. 1967, 5, 753–759.
[2] Zimm, B.H. J. Chem. Phys. 1948, 16, 1099–1116.
[3] Brandrup, J.; Immergut, E.H.; Grulke, E.A. (Eds.) Polymer Handbook, 4th ed.; Wiley: New York, 1999. (Mark-Houwink parameters)
[4] ASTM D6474-20 – Standard Test Method for Determining Molecular Weight Distribution and Molecular Weight Averages of Polyolefins by High Temperature GPC.
[5] ISO 16014-1:2019 – Plastics – Determination of average molecular weight and molecular weight distribution of polymers using size-exclusion chromatography.
[6] Polymer Char Application Notes and Publications Library. Available at: polymerchar.com/library/publications
[7] Modern Size-Exclusion Liquid Chromatography. Practice of Gel Permeation and Gel Filtration Chromatography (Wallace W. Yau, Joseph J. Kirkland, Donald D. Bly, Andre Striegel)
[8] Light Scattering, SEC and AF4: Powerful Tools for the Characterization of Polymers, Proteins and Nanoparticles (Stepan Podzimek)
To receive pricing or technical information about this instrument, complete the form below.